Medium Series Equipment

Ambient air sampling box for measuring pollutants (Fluorine & Chlorine) derived from UF6 near Uranium enrichment sites -
Critical operation 7 days a week, 24 hours a day - Network connection via SOFREL

Bubblers present upstream of the cabinet trap pollutants -
Vacuum pump creates a ΔP in order to sample the ambient air -
HR measurement associated with a membrane dryer protects the MFC -
Calculation of time of use / sampling allows to correlate the concentration of toxicant per m³ of air

Medium series equipment